Atlas FDA

Baxter Epiphany Cardio Server E3 ECG Management System Servers with software versions: a) v6.1.x, b) v6.2.x, c) v7.0.x.

FDA device recall Z-2647-2023 · posted 2023-09-28 · Open, Classified

Recall details

Recalling firm
Baxter Healthcare Corporation
Reason for recall
Epiphany, now a Baxter Healthcare company, is issuing an Urgent Medical Device Correction for the Cardio Server E3 ECG Management System software listed below due to a report of the caliper tool becoming non-responsive, preventing the caliper function from being used temporarily on software versions v6.1.x, v6.2.x, and v7.0.x of the E3 user interface.
Action taken
Baxter Healthcare initially notified consignees via email on 08/10/2023 of the issue and that a software patch would be implemented. An additional notification was sent to consignees via first-class mail on 08/30/2023. Consignees were instructed that Cardio Server support representatives will contact each facility to arrange for the installation of the software patch or consignees can proactively reach out to the Cardio Server Support team to schedule the patch update. Operators may continue to use the Cardio Server ECG Management System caliper tool prior to the installation of the software patch by following the steps listed in the notification letter. Customers were also instructed to notify other facilities or departments if the affected units have been further distributed, and to acknowledge receipt of the notification by responding on Baxter's portal at the website included in the letter.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
DQK
Quantity affected
89 units
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AL, AR, AZ, CA, CA, CT, FL, FL, GA, IA, ID, IL, IN, KS, KY, MA, MD, MO, MS, MT, NE, NH, NM, NV, NY, OH, OK, OR, P
Date initiated
2023-08-10
Date posted
2023-09-28
Location
Deerfield, IL

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
CARDIOSERVER (K052883)Datrix2005-11-01