Baxter Colleague Triple Channel Volumetric Infusion Pumps; Baxter Healthcare Corporation, Medication Delivery Division,

FDA device recall Z-1091-2007 · posted 2007-07-18 · Terminated

Recall details

Recalling firm
Baxter Healthcare Corp.
Reason for recall
A software anomaly is causing newly upgraded Colleague Triple Channel Infusion Pumps to alarm, display an error code (16:310:867:0002) and stop the infusion. This occurs during user programming with all three channels simultaneously infusing fluids. In reported cases, the pump stopped infusing and both an audible and a visual alarm notification were activated.
Action taken
Baxter telephoned the direct accounts on 6/20-21/07, informing them that Baxter has received multiple reports of newly upgraded Colleague Triple Channel Infusion Pumps alarming, displaying an error code and stopping infusion. These reported events occurred during user programming with all three channels simultaneously infusing fluids. In all cases, the pump stopped the infusions and both an audible and visual alarm notification were activated. The accounts were requested to immediately remove the affected triple channel pumps, identified by two attached hangtags entitled "Battery" and "Power On", from service. The accounts were informed that their Baxter representative would be in contact with them to coordinate the replacement of the affected pumps with upgraded single channel pumps or non-upgraded triple channel pumps, if available. Any questions were directed to Baxter's Medication Delivery Services at 1-800-843-7867. Baxter sent follow-up Urgent Product Recall letters dated 6/22/07.
Root cause
Software design
Status
Terminated
Product code
FRN
Quantity affected
4,512 pumps
Distribution
Worldwide, including USA, Canada, Australia, Belgium, New Zealand, South Africa, Thailand, and Singapore.
Date initiated
2007-06-20
Date posted
2007-07-18
Date terminated
2011-12-20
Location
Round Lake, IL