Baxter Colleague Single Channel Volumetric Infusion Pumps; Baxter Healthcare Corporation, I.V. Systems Division, Deerfie

FDA device recall Z-1022-05 · posted 2005-07-21 · Terminated

Recall details

Recalling firm
Baxter Healthcare Corp.
Reason for recall
The pumps may experience inadvertent power off, external communications port failures and electronic pump failure codes 402, 403, 533, 535, 599, 810:04 and 810:11.
Action taken
Baxter sent the 3/15/05 Urgent Device Correction letter to all Colleague Infusion Pump customers to alert them of actions they can take in the short-term to address specific user actions and several error codes in the event of an occurrence, which could lead to serious adverse health consequences if a patient is receiving life-sustaining medication at the time of the pump failure. Users were urged to verify that the pump is infusing after pressing the Start key. If the hospital uses the external monitoring capabilities of the pump for a nurse call or remote pump monitoring, they were requested to ensure that the external computer or monitoring system does not send data to the pump, and that the pump is powered off when connecting and disconnecting the cable to the DB9 port to avoid a 16:336 failure code. If a 16:336 failure occurs the user is advised to power the pump off and back on again. The users were advised to have a contigency plan to mitigate any disruptions of infusions of life sustaining drugs if failure codes beginning with 402, 403, 533, 535 and 599 occur, and to take the pump out of service for inspection by authorized service personnel. The users were also advised to avoid getting fluid on the tubing set or in the pump channel to lessen the occurence of the 810:04 and 810:11 failure codes. If those code failures occur, the pump should be taken out of service for inspection by authorized service personnel. The users were informed that software upgrades are being developed to correct the inadvertent power off problem and to mitigate failure code 16:336 occurrences. Any questions were directed to Baxter's Medication Delivery Services at 1-800-843-7867.
Root cause
Device Design
Status
Terminated
Product code
FRN
Quantity affected
152,260 pumps
Distribution
Nationwide including Puerto Rico, and internationally to Taiwan, Canada, Chile, Korea, Colombia, New Zealand, China, Hong Kong, Brazil, Turkey and Saudi Arabia.
Date initiated
2005-03-15
Date posted
2005-07-21
Date terminated
2011-06-01
Location
Round Lake, IL