Baxter Colleague, Colleague CX and Colleague CXE Infusion Pumps, Single Channel Volumetric Infusion Pumps. Baxter Health
FDA device recall Z-2222-2011 · posted 2011-05-19 · Terminated
Recall details
- Recalling firm
- Baxter Healthcare Corp.
- Reason for recall
- The Hi-Pot safety testing may not have been conducted on the affected pumps during service, prior to being returned to the end user.
- Action taken
- The firm, Baxter Healthcare Corporation, telephoned the customers on April 14, 2011, informing them that that the Hi-Pot test,and sent a follow up "URGENT DEVICE CORRECTION" letter dated April 27, 2011, to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to locate the pumps and take them out of service as soon as possible (Baxter will then send a shipping container for retrieval, re-inspection and return of the device back to the customer at no cost.); and return their colleague pumps for full functional testing including the HiPot test to the Baxter Buffalo Grove Global Technical Services (GTS) Center at Baxter GTS, 900 Corporate Grove Drive, Buffalo Grove, IL 60089. Baxter will perform full functional testing, including the HiPot test, on affected Colleague pumps for which the customer has continuing medical necessity. Should Baxter identify any issues through repair or service prior to returning the Colleague pumps to the customer. If you have questions regarding this communication, please call The Center for One Baxter at 1-800-422-9837 during the hours of 8:00 am to 5:00 pm CT, Monday through Friday.
- Root cause
- Employee error
- Status
- Terminated
- Product code
- FRN
- Quantity affected
- 117 pumps
- Distribution
- Worldwide distribution: USA (nationwide) including states of :Alabama, Arizona, California, Connecticut, Florida, Georgia, Illinois, Kentucky, Louisiana, Massachusetts, Maryland, Mississippi, North Ca
- Date initiated
- 2011-04-14
- Date posted
- 2011-05-19
- Date terminated
- 2012-05-09
- Location
- Round Lake, IL