Battery Pack portion of the LifeVest device. The LifeVest is a wearable cardioverter defibrillator worn by a patient at

FDA device recall Z-0061-2008 · posted 2007-10-18 · Terminated

Recall details

Recalling firm
Zoll Lifecor Corporation
Reason for recall
Possible malfunction of the battery pack. A significant electrostatic discharge (ESD) can cause the battery pack to malfunction. The malfunction causes the battery pack to discharge more rapidly and prevents the pack from recharging.
Action taken
Patients currently using the LifeVest should continue to do so while the product removal is underway. Advisory notices will be sent via first class mail 9/14/2007 to all active patients using the LifeVest device. Product replacements began on 8/31/07 by shipping out revised battery packs for new patient use. Shipments of new battery packs will continue based on production capacity and material availability. Any patients expected to be using the device longer than 5 months will be provided with 2 replacement Battery Packs. They will be instructed to return their existing packs in the same shipping box.
Root cause
Device Design
Status
Terminated
Product code
MVK
Quantity affected
2006 packs
Distribution
The product was shipped nationwide to patients. The product was also shipped to distributors in Germany, United Kingdom, Italy, Spain, France, and Denmark.
Date initiated
2007-09-14
Date posted
2007-10-18
Date terminated
2008-06-13
Location
Pittsburgh, PA