Atlas FDA

Basic Core Pack # OR1805, Sterile, Tri State Hospital Supply Corporation Manufacturers of Centurion Healthcare Products.

FDA device recall Z-1595-2010 · posted 2010-05-12 · Terminated

Recall details

Recalling firm
Centurion Medical Products
Reason for recall
The firm is conducting a sub recall of kits containing ConMed Goldline Rocker Switch Pencils. The pencils activation switch may remain in the "on" position after the switch has been depressed with excessive force, thereby continuously activating the pencil.
Action taken
Centurion sent out an "Urgent Product Recall Notice" to one customer on 4/9/2010. The customer was instructed to discontinue use of the product. The customer was also instructed to return any instock product .
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
GEI
Quantity affected
74 kits
Distribution
TX
Date initiated
2010-04-09
Date posted
2010-05-12
Date terminated
2010-10-25
Location
Howell, MI

Monitor recalls programmatically