Bard Mission Disposable Core Biopsy Instrument Kit: REF: 2010MSK, 20G x 10CM; REF: 2016MSK, 20G x 16CM

FDA device recall Z-1564-2023 · posted 2023-05-05 · Open, Classified

Recall details

Recalling firm
Bard Peripheral Vascular Inc
Reason for recall
Coaxial cannula internal diameter and length of biopsy instruments exceed the external diameter and length of the biopsy needle, prohibiting the biopsy needle from properly fitting into the coaxial cannula and accessing the target tissue, which could lead to procedure delay, and insufficient sample acquisition requiring a repeat procedure.
Action taken
On 04/11/23 recall notices were distributed to customers who were asked to do the following: 1. Discontinue use and dispose of affected product. 2. Share this notification with all users within your facility network of the product to ensure they are also aware of this field action. 3. If you purchased this product from a distributor, contact your distributor for further instructions and credit resolution or replacement product. 4. Complete and return the Customer Response Form to BRDC31@bd.com In addition, distributors were asked to do the following: Identify all customers within your distribution network that purchased any affected product as defined in this notification. Provide a copy of the attached customer letter to all customers to advise them of this field action notification. If you require further assistance please contact: North American Regional Complaint Center, Phone: 1-844-8BD- LIFE (1-844-823-5433), say "Recall" when prompted, M-F 8am - 5pm CT, or email: productcomplaints@bd.com
Root cause
Mixed-up of materials/components
Status
Open, Classified
Product code
KNW
Quantity affected
4,120
Distribution
US Nationwide Distribution: PR, AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD,
Date initiated
2023-04-11
Date posted
2023-05-05
Location
Tempe, AZ

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Bard Mission Disposable Core Biopsy Instrument (K171953)Bard Peripheral Vascular, Inc.2017-09-14