Atlas FDA

Bard Marquee Disposable Core Biopsy Instrument Kit: BARD MARQUEE KIT 12G X 10 CM, REF: MQK1210; BARD MARQUEE 12G X 10 CM

FDA device recall Z-0467-2025 · posted 2024-11-15 · Open, Classified

Recall details

Recalling firm
Bard Peripheral Vascular Inc
Reason for recall
Disposable biopsy instruments were potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device, which could lead to needle detachment and misplacement, which could lead to tissue damage and bleeding.
Action taken
On 9/30/24, removal notices were sent to customers who were asked to do the following: 1) Discontinue use of product listed in the letter, and immediately dispose of all affected product remaining in your possession in accordance with your local facility s process. 2) Share this notification with all users within your facility network of the product to ensure they are also aware of this field action. If the affected products have been distributed further, please share this notice to ensure that these entities are contacted and informed of the recall. 3) Complete and return the response form via email to BDRC27@BD.COM - If you purchased this product from a distributor, contact your distributor for further instructions and resolution. - Affected lots distributed from 05 April 2023 through 23 Aug 2024. If further assistance is required contact the firm: - North American Regional Complaint Center: 1-844-823-5433, Mon - Fri 8:00am and 5:00pm CT, productcomplaints@bd.com - Recall Related Questions: BDRC27@BD.COM On 1/15/2025, additional removal notices were mailed or emailed to distributors informing them of additional affected devices.
Root cause
Process change control
Status
Open, Classified
Product code
KNW
Quantity affected
71,498
Distribution
Worldwide - US Nationwide including in the states of WA, LA, MN, PA, UT, SC, OK, KY, AZ, IL, TX, NJ, NY, FL, TN, IN, OH, NC, CO, VA, GA, MD, CA, AL, WI, MO, IA, CT, NE, MI, MS, NV, KS, OR, SD, ID, NH,
Date initiated
2024-09-30
Date posted
2024-11-15
Location
Tempe, AZ

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
BARD(R) MONOPTY(R) DISPOSABLE CORE BIOPSY INSTRUMENT, BARD(R) MAX-CORE(R) DISPOSABLE CORE INSTRUMENT (K133948)Bard Peripheral Vascular, Inc.2014-02-21