Atlas FDA

BARD DORADO PTA BALLOON DILATATION CATHETERS, PRODUCT CODE # DR135412

FDA device recall Z-2122-2010 · posted 2010-08-02 · Terminated

Recall details

Recalling firm
Bard Peripheral Vascular Inc
Reason for recall
The affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft.
Action taken
The Bard Peripheral Vascular "URGENT MEDICAL DEVICE RECALL NOTIFICATION" with attached "RECALL AND EFFECTIVENESS CHECK FORM" was sent to the affected consignees beginning May 7, 2010 via FedEx. The recall notice informed the consignees of the reason for recall, the products affected and instructed them not to use or further distribute any affected product. The consignees were also instructed to call the firm's Recall Coordinator Silvia Carrillo at 1-800-321-4254 Option #2 Ext 2727 (M-F 7am to 4pm MST) or email at silvia.carrillo@crbard.com to facilitate the expedient return of the affected products. Additionally they were asked to fax the Recall and Effectiveness Check Form to BPV at 1-800-994-6772.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
DQY
Quantity affected
212 units
Distribution
Nationwide
Date initiated
2010-05-04
Date posted
2010-08-02
Date terminated
2012-02-01
Location
Tempe, AZ

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
DORADO PTA BALLOON DILATATION CATHETER (K072283)Bard Peripheral Vascular, Inc.2007-09-19