Bair Hugger Temperature Management Blanket, Model 555 Pediatric Full Access Blanket. Use only with a Bair Hugger tempera

FDA device recall Z-0855-05 · posted 2005-06-01 · Terminated

Recall details

Recalling firm
Arizant Inc
Reason for recall
Air channel seals on a limited number of Bair Hugger Model 555 Pediatric underbody blankets could separate when inflated with air from Bair Hugger warming unit. This separation would create a bulging area in the blanket's air channels, which could have a potential to shift the patient's position on the blanket.
Action taken
Recall notification was mailed to customers beginning 03/11/05. Customers were informed of the issue and asked to check their inventory for the affected lots. Affected lots found are to be quarantined and customers are to contact Arizant for instruction on returning affected product and receiving replacement product. Distributors are instructed to examine their inventory for affected lots, discontinue distribution and quarantine the lots. Distributors are asked to contact Arizant for returning affected product. If affected product was distributed, Distributors are to contact Arizant with names, addresses and telephone numbers of their accounts so Arizant can immediately inform those accounts of the recall.
Root cause
Other
Status
Terminated
Product code
DWJ
Quantity affected
88,320 blankets
Distribution
nationwide
Date initiated
2005-03-11
Date posted
2005-06-01
Date terminated
2007-01-10
Location
Eden Prairie, MN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
AUGUSTINE MEDICAL BAIR HUUGER, MODEL 750 TOTAL TEMPERATURE MANAGEMENT SYSTEM-(MODEL 750 WARMING UNIT, BAIR HUGGER BLANK (K001149)Augustine Medical, Inc.2000-09-06
BAIR HUGGER(R) PATIENT WARM SYST/BABY WARM COVER (K913734)Augustine Medical, Inc.1991-10-23