Atlas FDA

BacT/VIEW Software, all versions. in vitro diagnostic database software.

FDA device recall Z-0814-2007 · posted 2007-05-18 · Terminated

Recall details

Recalling firm
bioMerieux, Inc.
Reason for recall
BacT/VIEW will be unable to run scheduled macro applications unless the user touches the BacT/VIEW touch screen or clicks a function after resynchronizing the clocks.
Action taken
A customer notification was sent on/about December 19, 2006. Customers will be sent a new BacT/VIEW software version C.30b QTR 1 2007.
Root cause
Other
Status
Terminated
Product code
MDB
Quantity affected
1233 units
Distribution
Worldwide, including USA, Canada, UK, Italy, France, Spain, Germany, Poland, Chile, Japan and the Netherlands.
Date initiated
2006-12-19
Date posted
2007-05-18
Date terminated
2009-12-08
Location
Durham, NC