BacT/VIEW Software, all versions. in vitro diagnostic database software.
FDA device recall Z-0814-2007 · posted 2007-05-18 · Terminated
Recall details
- Recalling firm
- bioMerieux, Inc.
- Reason for recall
- BacT/VIEW will be unable to run scheduled macro applications unless the user touches the BacT/VIEW touch screen or clicks a function after resynchronizing the clocks.
- Action taken
- A customer notification was sent on/about December 19, 2006. Customers will be sent a new BacT/VIEW software version C.30b QTR 1 2007.
- Root cause
- Other
- Status
- Terminated
- Product code
- MDB
- Quantity affected
- 1233 units
- Distribution
- Worldwide, including USA, Canada, UK, Italy, France, Spain, Germany, Poland, Chile, Japan and the Netherlands.
- Date initiated
- 2006-12-19
- Date posted
- 2007-05-18
- Date terminated
- 2009-12-08
- Location
- Durham, NC