Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063, 722221), Azurion 3 M15 (722064, 722222),

FDA device recall Z-2258-2026 · posted 2026-05-28 · Open, Classified

Recall details

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Reason for recall
It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.
Action taken
On April 30, 2026, URGENT Medical Device Correction letters were sent to customers. Actions to be taken: 1. Circulate this URGENT Medical Device Correction letter to all users of the system so that they are aware of the issue. 2. In case motorized movements are unavailable, the stand can be manually repositioned using the handgrips and brake controls located on both sides of the stand. For ceiling-mounted systems, manual longitudinal (1), transversal (2) and rotational (3) stand movements remain available. 3. In case the affected system has been transferred to another organization, please send a copy of this URGENT Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. 4. Keep this URGENT Medical Device Correction letter with the documentation of the system until Philips corrects your system. Ensure that the letter is in a place likely to be seen/viewed. 5. If you experience the issue described in this letter, please report it to your local Philips representative. Actions by Philips: Philips will replace the HDDs in all affected systems. Philips expects to start replacing the HDDs in September 2026. A local Philips representative will contact you to schedule a visit to replace the HDDs in your system.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
OWB
Quantity affected
3069 units (872 US, 2224 OUS)
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Afghanistan, Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina
Date initiated
2026-04-30
Date posted
2026-05-28
Location
Best, Netherlands

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Allura Xper FD series and Allura Xper OR Table series (K161563)Philips Medical Systems Nederland B.V.2016-07-29
ALLURA XPER FD SERIES / ALLURA XPER OR TABLE SERIES (K141979)Philips Medical Systems Netherland BV2014-08-19
ALLURA XPER FD; OR TABLE (K133292)Philips Medical Systems Nederland B.V.2014-03-05
ALLURACLARITY XPER FD SERIES X-RAY SYSTEM (K130638)Philips Medical Systems Nederland B.V.2013-06-28
ALLURA XPER FD OR TABLE SERIES; ALLURA XPER FD10 SERIES; ALLURA XPER FD20 SERIES; ALLURA XPER FD BIPLANE SERIES (K130842)Philips Medical Systems Nederland B.V.2013-05-21
ALLURA XPER OR TABLE SERIES (K102005)Philips Medical Systems North America, Inc.2010-08-09