Atlas FDA

Azurion 3 M12, Azurion 3 M15, Allura R9 7 M12, Allura R9 M20, Azurion 7 M12, Azurion 7 M20 The Allura Xper series is int

FDA device recall Z-0255-2018 · posted 2017-12-18 · Terminated

Recall details

Recalling firm
Philips Electronics North America Corporation
Reason for recall
When using the downscale option with the Allura R9 system or Azurion R1.1 system, the measurements performed using the QA Basic Measurement tool will not be correctly exported to the external DICOM destination.
Action taken
Image Guided Therapy Systems ( IGT) sent an Urgent - Field Safety Notice Medical Device Correction letter dated August 11. 2017. The Downscale option should be disabled in all Export protocols until Philips corrects this issue. To accomplish this, you may refer to Chapter 13.8 (Configuring Export Protocols) of the system instructions for use that describes how to perform this action. You may also contact your Philips local representative for assistance in performing this action. All archived images exported using the downscale option would be affected by this issue, therefore such measurements from archived images should not be relied on. Additional precaution could be taken by re-performing measurements during the actual interventional procedure. The problem will be resolved by a software update, which is expected to be available by the end of 2017. In the interim, a similar notice will be provided along with newly manufactured devices shipped for use. For further questions, please call (800) 722-9377.
Root cause
Software design
Status
Terminated
Product code
OWB
Quantity affected
157
Distribution
Worldwide Distribution - USA (nationwide) Distribution.
Date initiated
2017-08-18
Date posted
2017-12-18
Date terminated
2019-09-26
Location
Andover, MA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Allura Xper R9 (K162148)Philips Medical Systems Nethrlands BV2016-11-23