AxSYM CORE 2.0 Reagent Pack; Abbott Laboratories, Abbott Park, IL 60064; AxSYM CORE 2.0 is a microparticle enzyme immuno
FDA device recall Z-1048-2009 · posted 2009-04-15 · Terminated
Recall details
- Recalling firm
- Abbott Laboratories
- Reason for recall
- Defective caps: Customers have observed reagent bottles not opening appropriately, which could lead to a probe crash on the instrument.
- Action taken
- Consignees were sent a Product Correction-Immediate Action Required letter on 1/19/09. The letter informed users of the reported increased frequency of probe crashes when using the affected reagent pack lots and provided instructions on what actions the laboratory can take if they have a defective reagent pack or experience a probe crash with their system. Users need to inspect each reagent pack for broken or improperly sealed caps; manually open each bottle to ensure the flipper bar raises the reagent cap correctly; close each flipper bar prior to loading reagent pack on instrument; and discard any reagent packs that do not open properly. If a probe crash is observed prior to testing samples again, a probe crash recovery must be performed following the procedure in the Operations Manual. A copy of the communication needs to be provided to any laboratory that you may have forwarded affected lots to. Users were requested to return a customer reply letter acknowledging receipt and understanding of the letter. For additional information, contact Abbott Labs at 1-877-4ABBOTT.
- Root cause
- Employee error
- Status
- Terminated
- Product code
- LOM
- Quantity affected
- 967 kits
- Distribution
- Worldwide Distribution -- USA including Puerto Rico and countries of Germany, Mexico, Hong Kong, and Japan.
- Date initiated
- 2009-01-19
- Date posted
- 2009-04-15
- Date terminated
- 2010-09-11
- Location
- Abbott Park, IL