AXIOM SIRESKOP SD, SIRESKOP SX, AXIOM LUMINOS System, X-Ray, Fluoroscopic, image-intensified; solid state x-ray imager (

FDA device recall Z-1421-2011 · posted 2011-02-23 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
Firm became aware of a potential malfunction and possible hazard to patients in the event the tilt buttons on the OPTI Grip become stuck under the housing
Action taken
Siemens Medical Solutions USA, Inc issued a Customer Safety Advisory Notice via Update Instruction XP005/10/S to affected customers. This letter informs customers of the potential malfunction and of possible danger for patients, users or other persons. A future Update Instruction will be released for the replacement of the tilt buttons on all OPTI Grip handles. Customers were instructed to forward the safety notice to any new owners of this device and to provide Siemens Medical with names of the new customers. For any questions regarding this recall call 610-448-3237.
Root cause
Other
Status
Terminated
Product code
MQB
Quantity affected
1323
Distribution
Nationwide Distribution including AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN
Date initiated
2011-01-14
Date posted
2011-02-23
Date terminated
2013-07-16
Location
Malvern, PA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
AXIOM LUMINOS DRF (K062623)Siemens Medical Solutions USA, Inc.2007-08-22
AXIOM SIRESKOP SD (K051602)Siemens Medical Systems, Inc.2005-07-07
SIRESKOP SX (K951358)Siemens Medical Solutions USA, Inc.1995-05-15