AXIOM Artis or Artis zee system Product Usage: Axiom Artis and Artis zee / zeego are dedicated angiography systems devel

FDA device recall Z-2587-2014 · posted 2014-09-24 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
There is a potential problem with the AXIOM Artis or Artis zee system in connection with the hand switch operated at the patient table which may result in uncontrolled release of an x-ray image.
Action taken
Siemens sent an Important Safety Information letter dated June 5, 2014, to affected customers. The letter described the affected product, problem, and actions to be taken. Customers were instructed to observe this safety notice and comply with the corresponding measures until the update has been fully completed.
Root cause
Device Design
Status
Terminated
Product code
IZI
Quantity affected
1067
Distribution
US Nationwide Distribution.
Date initiated
2014-06-05
Date posted
2014-09-24
Date terminated
2015-04-06
Location
Malvern, PA