Aviator Drill Bit (10 mm, 12 mm, 14 mm, 16 mm, Manufactured by: Stryker Spine SAS Z.1 Marticot -33610- Cestas France +33

FDA device recall Z-0163-2013 · posted 2012-10-29 · Terminated

Recall details

Recalling firm
Stryker Spine
Reason for recall
Aviator drill bits are 10 mm longer than the Reflex Hybrid drill bits and are not interchangeable which could lead to over-drilling by as much as 10 mm.
Action taken
Stryker Spine sent an Urgent Medical Device Field Correction Notification letter beginning July 5, 2012, to all affected customers. Teh letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check the drill bit length in each respective drill guide prior to surgery to ensure the correct length protrudes through the guide. Customers with questions should call 201-760-8298. Customers were asked to report any adverse events or product quality problems to 1-888-457-7463.
Root cause
Device Design
Status
Terminated
Product code
LXH
Quantity affected
6154 units
Distribution
Nationwide distribution.
Date initiated
2012-07-05
Date posted
2012-10-29
Date terminated
2013-01-28
Location
Allendale, NJ