Atlas FDA

Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A138 used in conjunction with Warming

FDA device recall Z-1172-2025 · posted 2025-02-14 · Open, Classified

Recall details

Recalling firm
Augustine Temperature Management, LLC
Reason for recall
Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact .
Action taken
Augustine Temperature Management contacted their consignees by visit beginning on 02/15/2024 to replace the affected cables. The firm collected email confirmation that the original cables had either been returned or destroyed. If you have any questions, contact the Product Manager at 952-465-3526 or email:
Root cause
Device Design
Status
Open, Classified
Product code
DWJ
Quantity affected
7 unit
Distribution
US distribution to states of: AL, TX, NV
Date initiated
2024-02-15
Date posted
2025-02-14
Location
Minnetonka, MN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
HotDog Warming Mattress + Return Electrode (K230866)Augustine Temperature Management, LLC2023-06-29
HotDog Warming Mattress + Return Electrode (K210727)Augustine Temperature Management, LLC2022-02-01