ATTUNE FB PS ARTICULATION SURFACE SIZES 3-8 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the

FDA device recall Z-2137-2015 · posted 2015-07-20 · Terminated

Recall details

Recalling firm
DePuy Orthopaedics, Inc.
Reason for recall
Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.
Action taken
URGENT INFORMATION: MEDICAL DEVICE CORRECTION was issued for all lots of the affected devices. The device correction is to inform US Distributors and Medical Professionals about the issue, how to prevent the issue, and any actions to take. US Distributors will provide Hospitals/User Facilities with the Medical Professional Notice and Acknowledgment Cards. US Distributors will also work with the sales consultants to ensure all consultants complete sales training regarding this issue.
Root cause
Device Design
Status
Terminated
Product code
MBH
Quantity affected
34676 units
Distribution
Worldwide Distribution - US including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD
Date initiated
2015-06-12
Date posted
2015-07-20
Date terminated
2015-11-12
Location
Warsaw, IN