ATAR Reusable Extension Cable, sold under brand names: ATAR Reusable Series, ATAR 53912 Reusable Series, and XI Reusable
FDA device recall Z-2340-2017 · posted 2017-04-10 · Terminated
Recall details
- Recalling firm
- Oscor, Inc.
- Reason for recall
- The Reusable Oscor ATAR extension cables could separate from the connector during use.
- Action taken
- Oscor sent an Important Information letter dated March 27, 2017, to all affected customers. The letter explained that they awee aware the Reusable Oscor ATAR extension cables could separate from the connector during use when instructions for use were not followed. They pointed out to consignees that if reused, they can only be reused a maximum of 2 times and must be resterilized by Oscor EtO. Customers were notified again, when the recall expanded on 06/01/2017, and were sent notifications on 06/05/2017. Instructions to customers included to examine inventory, quarantine any affected devices, notify customers if further distributed, and contact Oscor to arrange for replacement. Customers were notified a third time when the recall expanded on 06/13/2017, and notifications were sent between 06/23/2017 and 07/01/2017. Instructions included to check inventory, quarantine any affected devices, return the Acknowledgement Form, and return affected cables to Oscor. Customers with questions should call 727-937-2511.
- Root cause
- Process control
- Status
- Terminated
- Product code
- IKD
- Quantity affected
- 19,178 units
- Distribution
- Worldwide Distribution -US to AL, AK, CA, CO, FL, GA, ID, IL, IN, IA , KS, KY, ME, MD, MA, MI, MN, MO, MT, NE, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WV, WI, DC. Foreign distr
- Date initiated
- 2017-03-31
- Date posted
- 2017-04-10
- Date terminated
- 2024-07-15
- Location
- Palm Harbor, FL
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| ELECTRODE/EXTENSION CABLES, MODELS ATAR (K070926) | Oscor, Inc. | 2007-06-07 |