ASP-1020 LipoTower with Aspiration Pump. Medical use.
FDA device recall Z-2821-2015 · posted 2015-09-30 · Terminated
Recall details
- Recalling firm
- MicroAire Surgical Instruments, LLC
- Reason for recall
- There is potential risk of electrical shock if lead wires inside the device become detached and the ground connection fails.
- Action taken
- The firm has issued FDA recall letters to all consignees. The letter have been sent out by UPS to expedite the delivery.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- MUU
- Quantity affected
- 3 LipoTower with Aspiration Pump
- Distribution
- Nationwide Distribution.
- Date initiated
- 2015-06-12
- Date posted
- 2015-09-30
- Date terminated
- 2016-05-23
- Location
- Charlottesville, VA