Aruba Wireless Network Access Points, Model 65 utilized by the Acuity Central Monitoring System. Welch Allyn Protocol, I

FDA device recall Z-2918-2011 · posted 2011-08-03 · Terminated

Recall details

Recalling firm
Welch Allyn Protocol, Inc
Reason for recall
Aruba Wireless Network Access Points, Model Number 65, experienced a higher than expected incidence of dropouts due to external RF interference.
Action taken
Welch Allyn Protocol, Inc. sent a letter dated May 16, 2011 to affected customer. The letter stated there were concerns over drop-outs in the system recommended the system at this hospital to revert back to the Symbol wireless infrastructure. For questions call Welch Allyn at 315-685-4100.
Root cause
Software design
Status
Terminated
Product code
DSI
Quantity affected
12 units were affected
Distribution
Worldwide distribution: (USA) Nationwide distribution including the state of California; and the countries of Singapore; Philippines, France, Italy, Malaysia, Poland, UK, England, UAE, Germany, Spain,
Date initiated
2011-05-16
Date posted
2011-08-03
Date terminated
2011-08-12
Location
Beaverton, OR

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ACUITY CENTRAL STATION, MODEL 020XXXXX (NOTE: XXXXX = VARIOUS CONFIG.) (K052160)Welch Allyn, Inc.2005-12-16