ARTISTE, ONCOR Avantgarde, ONCOR Impression Plus, ONCOR Expression, PRIMUS, PRIMUS Plus, Syngo RT Therapist; therapeutic
FDA device recall Z-2812-2015 · posted 2015-09-24 · Terminated
Recall details
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Reason for recall
- A software fix has been released to prevent automatic movement resulting in a collision safety risk for patients.
- Action taken
- Siemens sent Notification letters to affected customers. The letter identified the affected product, problem and actions to be taken. The letter advise customers that the software fix was rolled out to address multiple safety issues.
- Root cause
- Software design (manufacturing process)
- Status
- Terminated
- Product code
- IYE
- Quantity affected
- 26
- Distribution
- Worldwide Distribution and US Nationwide
- Date initiated
- 2015-08-05
- Date posted
- 2015-09-24
- Date terminated
- 2016-09-01
- Location
- Malvern, PA