Atlas FDA

ARTISTE , ONCOR and PRIMUS accelerators that utilize the RT Therapist v4.2 system in combination with syngo RT Oncologis

FDA device recall Z-2216-2011 · posted 2011-05-19 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
Siemens has become aware of a potential safety issue that may result in mistreatment when customers are configured with the ARTISTE Linac RT Therapist v4.2 in combination with RT Oncologist v.4.2 and use the remote "Online 2D Image Review.
Action taken
The firm, Siemens, sent an "Urgent: Medical Device Correction Customer Safety Advisory Notice" on March 31, 2011 via Siemens Customer Service Group or certified mail to its customers. The notice describes the product, problem and actions to be taken. The customers were instructed to not use the "Online 2D Image Review" on the Syngo RT Oncologist 4.2; to always calculate the offset directly on the RTT 4.2 before it is applied for the treatment with the patient on the treatment table; include this notice in their syngo RT Therapist 4.2 System Owner Manual chapter "Safety Advisory Letter" where it should remain until the measures defined in the notice are performed, and in the interest of safety, perform the preventative measures noted in this notice and inform all affected personnel immediately. Siemens is currently working on a solution for the incremental loading workflow in syngo RT Therapist 4.2 and will provide a software update as soon as possible. Siemens will notify the customers as soon as an update for their system will be available. UPDATE: The release of Safety Update UI TH010/11/S to distribute the software update for RTTv4.2.94O04 began on April 15, 2011. If you have any questions, contact Regulatory Affairs at 925-602-8157.
Root cause
Software design
Status
Terminated
Product code
IYE
Quantity affected
45 active devices
Distribution
Worldwide distribution: USA (nationwide) including states of: CA, FL, OH, PA and UT; and countries including: Australia, Belgium, Croatia, Germany, Japan, South Africa, Sweden and United Kingdom.
Date initiated
2011-03-31
Date posted
2011-05-19
Date terminated
2012-03-30
Location
Concord, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
MODIFICATION TO HEMCON BANDAGE AND HEMCON BANDAGE OTC (K072486)Hem Con Medical Technologies, Inc.2008-08-06
ONCOR EXPRESSION (K060226)Siemens Medical Solutions USA, Inc.2006-03-15
PRIMUS (K993425)Siemens Medical Solutions USA, Inc.1999-11-26