Artiste MV Digital Linear Accelerator, when used in combination with Syngo RT Therapist Version 4.1 build 67 and automat
FDA device recall Z-0857-2009 · posted 2009-01-14 · Terminated
Recall details
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Reason for recall
- Under certain conditions, the flat panel may move unexpectedly and collide with the patient.
- Action taken
- The firm issued an Urgent Medical Device Customer Safety Letter, dated 10/23/2008. With the letter included a detailed description of the issue, the condition under which this issue may occur, and recommendation to avoid them. Customers are to read the information carefully and distribute to the responsible persons. The Customer Safety Letter should be included their syngo RT Therapist Express manual and remain there until the next release of software is installed. In the interest of safety, the preventive measures should be performed and all affected personnel should be informed immediately. The firm is tracking responses via return cards and completion protocol.
- Root cause
- Labeling design
- Status
- Terminated
- Product code
- IYE
- Quantity affected
- 31 consignees have the configuration noted above.
- Distribution
- Worldwide Distribution --- including USA and countries of Netherlands, Germany, Poland, Norway, Australia, Belgium, France, New Zealand, Malaysia, Spain and Canada.
- Date initiated
- 2008-10-23
- Date posted
- 2009-01-14
- Date terminated
- 2010-12-09
- Location
- Concord, CA