Artiste Maverick Tint, Part Number: N36TB, Lot/Serial Numbers: 3416180, Product Description and Intended Use: The intend
FDA device recall Z-0455-2012 · posted 2012-01-11 · Terminated
Recall details
- Recalling firm
- Kerr Corporation
- Reason for recall
- The resin material separates when the product is extruded from the syringe. Although there are no safety issues related with this situation, please refrain from using the affected product.
- Action taken
- Pentron Clinical sent an URGENT: Medical Device Recall letter dated April 20, 2011 to all affected customers. The letter described the product, problem and actions to be taken. Customers were instructed to examine their inventory and quarantine all product(s) subject to recall for return and exchange at no charge. In addition, the letter states that if the product was further distributed to contact their customers and notify them of the recall. Customers were also asked to complete and return the enclosed Recall Return Form. Contact Pentron Clinical Customer Service at (800) 551-0283, (option 1) directly to handle the arrangement of a quick return and replacement.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- EMA
- Quantity affected
- 100 Units
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of: Armenia, Russia and Ukraine.
- Date initiated
- 2011-03-03
- Date posted
- 2012-01-11
- Date terminated
- 2012-01-09
- Location
- Orange, CA