Artis Systems with Software Artis VC20x/VC21A/VD10x that have a DSA license installed. Angiographic x-ray system

FDA device recall Z-0997-2013 · posted 2013-03-26 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
The firm has become aware of a potential issue on Artis systems with the software Artis VC20x/VC21A/VD10x that have a DSA license installed. Under certain preconditions, executing roadmap OGP (organ program) in the DSA overlay mode may lead to imprecise registration of the DSA mask image with the roadmap image displayed on the live monitor.
Action taken
Siemens sent a Customer Safety Advisory Notice on March 8, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to pass the information on to all those who need to be aware of it and to instruct their personnel accordingly. Customers were asked to forward the safety information to other organizations that couald be affected by this action. For questions regarding this recall call 610-219-6300.
Root cause
Software design
Status
Terminated
Product code
IZI
Quantity affected
81
Distribution
Nationwide Distribution including AL, CA, CO, FL, IA, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, ND, NJ, NY, OH, PA, SC, TX, UT, VA, WA, and WI and to Puerto Rico.
Date initiated
2013-03-08
Date posted
2013-03-26
Date terminated
2014-11-06
Location
Malvern, PA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ARTIS ZEE / ZEEGOS WITH SW VC14 (K090745)Siemens Medical Solutions USA, Inc.2009-06-18
ARTIS ZEE AND ZEEGO (K073290)Siemens Medical Solutions USA, Inc.2008-02-11