ARTIS pheno, Interventional fluoroscopic x-ray system, Model No. 10849000
FDA device recall Z-0119-2023 · posted 2022-10-21 · Open, Classified
Recall details
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Reason for recall
- Siemens has become aware of three potential software issues with ARTIS pheno systems with software version VE10B. This may lead to a hazardous situation for patients if treatment cannot be continued on the system and treatment needs to be continued on an alternate system.
- Action taken
- On September 20, 2022, the firm notified affected customers via Urgent Medical Device Correction letters. This notification covered three issues. 1. System Error Management, in which the operator is unable to release X-Ray in "Bypass fluoroscopy" mode which remains permanently inhibited, and may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system. Customers should ensure that patient treatment can be continued in other ways if there is any possible danger for the safety of the patient. 2. Head Holder Symbols, in which the user cannot tell whether a head holder has been selected in the system, may lead to a collision between the C-Arm and the head holder during system movement if the head holder is being used but not selected on the system. Customers should ensure that the configuration of the head holder is correct prior to system movements and double check configuration in the control room in case of uncertainty during the procedure. 3. Stand Movement, in which rarely no stand movement is possible after a routine brake test, leading to a possible delay of procedure. Customers should carry out routine brake tests with sufficient time before starting any procedure. Siemens will correct the software errors and contact customers to arrange a date to perform the corrective action. Please feel free to contact Siemens' service organization for an earlier appointment at 1-800-888-7436.
- Root cause
- Software design
- Status
- Open, Classified
- Product code
- JAA
- Quantity affected
- 140 US
- Distribution
- Domestic distribution nationwide. Foreign distribution worldwide.
- Date initiated
- 2022-09-20
- Date posted
- 2022-10-21
- Location
- Malvern, PA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| ARTIS pheno (K163286) | Siemens Medical Solution USA, Inc. | 2017-03-09 |