ARROW PICC Kit featuring Chlorag+ard Technology Midline Catheterization Kit ARROW Advantage5 Pressure Injectable PICC Ki
FDA device recall Z-0207-2016 · posted 2015-11-04 · Terminated
Recall details
- Recalling firm
- Arrow International Inc
- Reason for recall
- Possibility that the catheter peel-away component hub tabs may prematurely detach when the practitioner begins to peel apart the sheath body from the catheter. Although the issue does not pose a significant risk to the patient, it may cause a delay of treatment when placing the catheter.
- Action taken
- Arrow International sent an Urgent Medical Device Recall Notification letter dated October 16, 2015, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignees. Consignees were instructed to immediately discontinue use and quarantine any effected products. Consignees were instructed to return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507 Attn: Customer Service or email it to recalls@teleflex.com. A customer service representative will contact consignees with a Return Goods Authorization (RGA) Number and will provide instructions for the return of product to Arrow International. For consignees with no affected product, should also complete the enclosed Recall Acknowledgement Form. For questions customers should call 1-866-246-6990.
- Root cause
- Unknown/Undetermined by firm
- Status
- Terminated
- Product code
- LJS
- Quantity affected
- 12,157
- Distribution
- World Wide Distribution - US including AR, AL, TX, MA, CA, IA, MO, NY, AL, VA, RI, IN, TN, FL, KY, NH, SC, MD, WA, OK, IL, ME, OR, OH, PA, MI, KS, LA, TN, WV, NE, NJ, NV, GA, SD, CT, MS, CO, and Inter
- Date initiated
- 2015-10-16
- Date posted
- 2015-11-04
- Date terminated
- 2017-11-09
- Location
- Reading, PA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| CG+ ARROW JACC POWERED BY ARROW VPS STYLET (K132133) | Arrow International (Subsidiary of Teleflex Inc.) | 2013-09-26 |
| ARROW GLIDETHRU PEEL-AWAY SHEATH/DILATOR INTRODUCER (K122854) | Arrow International (Subsidiary of Teleflex Inc.) | 2013-01-04 |
| ARROW PICC POWERED BY ARROW VPS STYLET (K121941) | Arrow International (Subsidiary of Teleflex Inc.) | 2012-09-07 |
| ARROWADVANTAGES PRESSURE INJECTABLE PERIPHERALLY INSCRTED CENTRAL CATHETER KITS (K113277) | Arrow International, Inc. | 2012-06-21 |
| ARRW EVOLUTION (K112896) | Arrow International (Subsidiary of Teleflex Inc.) | 2012-05-10 |