Arrow EVOLUTION Pressure Injectable PICC with Chlorag+ard Antimicrobial and Antithrombogenic Technology

FDA device recall Z-2698-2017 · posted 2017-06-27 · Terminated

Recall details

Recalling firm
Arrow International Inc
Reason for recall
Arrow is recalling additional lots that were identified as part of an active recall. Arrow is recalling the affected product due to the possibility that the catheter peel-away component hub tabs may prematurely detach when the practitioner begins to peel apart the sheath body from the catheter.
Action taken
Teleflex/Arrow International mailed an Urgent Medical Device Recall Notification Letter to affected customers on 06/12/2017 to inform them of the issue. Arrow requested that customers examine their inventory immediately for the affected lots and discontinue use and quarantine any products with the associated product codes identified in the notice and complete the Recall Acknowledgement Form and fax back to the number included in the notice.
Root cause
Device Design
Status
Terminated
Product code
LJS
Quantity affected
9,037 units in the U.S. and 4,500 units internationally
Distribution
Distributed to SC, AL, NJ, IN, MA, GA, CA, PA, AZ, VA , WA and Bangkok
Date initiated
2017-06-12
Date posted
2017-06-27
Date terminated
2018-09-06
Location
Reading, PA