arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Urea

FDA device recall Z-0631-2016 · posted 2016-01-15 · Terminated

Recall details

Recalling firm
Arkray Factory USA, Inc.
Reason for recall
One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL.
Action taken
The firm sent an Urgent Medical Device Recall letter to customers beginning 12/18/2015. The letter identified affected product, stated the reason for recall and asked for unused product to be returned. For instructions on returning unused product contact Lesty Jaranilla 1-877-538-8872 x3243.
Root cause
Component design/selection
Status
Terminated
Product code
CGA
Quantity affected
11 boxes (25 foil packaged test strips per box)
Distribution
Distributed in the states of FL, IL, KY, MI, NC, NY, OH, and TN.
Date initiated
2015-12-18
Date posted
2016-01-15
Date terminated
2017-01-04
Location
Minneapolis, MN