ARCHITECT Total T3 Reagent Kit (7K64) consists of: - 1 or 4 Bottle(s) (6.6 mL/27.0 mL) anti-T3 (sheep) coated microparti

FDA device recall Z-0474-2015 · posted 2014-12-03 · Terminated

Recall details

Recalling firm
Abbott Ireland Diagnostics Division
Reason for recall
17% of ARCHITECT Total T3 Reagent Kit, LN 7K64-20/-25 reagent lots 38901UI00 and 38901UI01 may exhibit lower Relative Light Units (RLUs) than expected, which in turn may result in controls out of range or patient results higher than expected.
Action taken
Abbott sent an Product Recall letter dated September 11, 2014, to all direct accounts (customers). The letters included instructions for customers to: 1) Discontinue use of and destroy any remaining inventory according to your laboratory procedures; 2) follow your laboratory protocol regarding the need for review of previously reported patient results; 3) replacement material is available; for ordering, please order through your normal ordering process; and, 4) if you have forwarded any of the affected lots to another laboratory, please provide them with a copy of this communication. Customers or healthcare providers with questions can contact Customer Service at 1-877-422-2688 (24 hours a day, 7 days a week).
Root cause
Employee error
Status
Terminated
Product code
CDP
Quantity affected
7,415 kits
Distribution
Worldwide Distribution - USA (nationwide) including Puerto Rico and the states of AL , AR , AZ , CA , CT , FL , GA , IL , IN , KS , LA , MA , MD , MI , MN , MO , MS , MT , NC , NJ , NV , NY , OH , OK
Date initiated
2014-09-11
Date posted
2014-12-03
Date terminated
2016-12-23
Location
Lisnamuck Co. Longford, Ireland