Atlas FDA

ARCHITECT STAT Troponin-I Reagent Kit, list 2K41-20; each 400 test kit contains 4 bottles each of microparticles, conjug

FDA device recall Z-0513-06 · posted 2006-02-14 · Terminated

Recall details

Recalling firm
Abbott Laboratories MPG
Reason for recall
There has been an increase in customer complaints observed for failure to calibrate or inability to generate patient result errors.
Action taken
Abbott sent recall letters dated 12/20/05 to the laboratory accounts via Federal Express Priority on 12/21/05. The accounts were informed of the increase in customer complaints for failure to calibrate or inability to generate patient result errors for the Troponin-I Reagent lot, and were requested to discontinue use and destroy any of the affected lot found in inventory for replacement.
Root cause
Other
Status
Terminated
Product code
MMI
Quantity affected
181 kits
Distribution
Nationwide and internationally to Canada, Australia, Germany and Taiwan
Date initiated
2005-12-21
Date posted
2006-02-14
Date terminated
2006-07-31
Location
Abbott Park, IL