Architect Sirolimus Reagent Pack; 100 test pack containing 1 bottle (8.0 mL) Anti-Sirolimus Coated Microparticles, 1 bot

FDA device recall Z-2172-2010 · posted 2010-08-03 · Terminated

Recall details

Recalling firm
Abbott Laboratories
Reason for recall
The Architect Sirolimus assay lot 80162M100 is exhibiting higher than normal reports of barcode read errors.
Action taken
Abbott sent a "Product Recall" letter dated May 27, 2010, to all customers receiving the affected lot of reagent. The customers were informed of the increased frequency of barcode read errors due to the print quality of the barcode label. The following error codes have been observed for some kits within the lot: 4000 - Unable to read reagent bar code in position (x) on (y) carousel, 0208 - Reagent carousel scan error, and 2011 - Extra reagent bottle detected in position (x) on (y) carousel. The customers were requested to determine if they are currently using or have inventory of lot 80162M100, discontinue use of and destroy any remaining inventory of the lot, and retain a copy of the letter for their laboratory records. Customers requiring additional information were directed to contact Abbott Customer Service at 1-877-422-2688. Customers outside the U.S., please contact your local Customer Service representative.
Root cause
Equipment maintenance
Status
Terminated
Product code
NRP
Quantity affected
875 kits
Distribution
Worldwide distribution: USA including Puerto Rico, and countries of Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Egypt, Estonia, Finland, Fran
Date initiated
2010-05-27
Date posted
2010-08-03
Date terminated
2011-07-15
Location
Abbott Park, IL