ARCHITECT Lactic Acid Part # 09D891T21 The Lactic Acid assay is used for the quantitation of lactic acid in human plasma
FDA device recall Z-0379-2017 · posted 2016-11-09 · Terminated
Recall details
- Recalling firm
- Abbott Laboratories
- Reason for recall
- Abbott has identified negative interference from the drug N-Acetyl Cysteine (NAC) with the ARCHITECT Lactic Acid reagent (LN 9D89-21).
- Action taken
- Abbott Laboratories sent an Product Correction Immediate Action Required letter dated October 10, 2016, to all affected customers. Please review this letter with your Medical Director Taking the above information into consideration, you may continue to use ARCHITECT Lactic Acid reagent (LN 9D89-21) Please complete and return the Customer Reply Form If you have forwarded the product listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter Please retain this letter for your laboratory records. If you or any of the health care providers you serve have any questions regarding this information, U.S. Customers please contact Customer Service at 1-877-4ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., please contact your local area Customer Service. For further questions please call (847) 937-2550.
- Root cause
- Under Investigation by firm
- Status
- Terminated
- Product code
- KHP
- Quantity affected
- 26, 837 kits
- Distribution
- Nationwide and Worldwide Distribution.
- Date initiated
- 2016-10-11
- Date posted
- 2016-11-09
- Date terminated
- 2017-05-17
- Location
- Abbott Park, IL