ARCHITECT Free T4 Calibrator Kit, list 7G96-01; the kit consists of 2 bottles of calibrators, CAL 1 - 0.5 and CAL 2 - 6.

FDA device recall Z-0353-04 · posted 2004-07-20 · Terminated

Recall details

Recalling firm
Abbott Laboratories HPD/ADD/GPRD
Reason for recall
The calibrator lots may yield Free T4 values that are depressed.
Action taken
Architect Free T4 Calibrator kits were recalled by letter dated 12/15/03. The accounts were informed of the out of range control values, and the potential for depressed patient values with the kits, and were requested to discontinue use and destroy the affected lots of product.
Root cause
Other
Status
Terminated
Product code
CEC
Quantity affected
1,364 kits
Distribution
Nationwide and internationally to Abbott subsidiaries in Brazil, Canada, Germany, Japan, Singapore, Hong Kong, Thailand, Korea, Australia, Colombia, New Zealand, Venezuela and Taiwan.
Date initiated
2003-12-15
Date posted
2004-07-20
Date terminated
2004-07-22
Location
Abbott Park, IL