ARCHITECT Folate Reagent Kit, list 6C12-20 ( 4 x 100 tests), 6C12-25 (1 x 100 tests) and 6C12-30 (4 x 500 tests), Abbott

FDA device recall Z-0783-05 · posted 2005-04-28 · Terminated

Recall details

Recalling firm
Abbott Laboratories MPG
Reason for recall
concentrations for normal samples generated with current reagent lots of the ARCHITECT Folate assay are lower than data generated for the package insert.
Action taken
Abbott international affilates were notified of the recall by e-mail on 03/11/2005, and requested to notifiy their customers. The letters inform the end users of the downward shift in control and/or patient results, and requested them to assess their normal and deficient ranges for their specific population, assess their quality control ranges and follow their laboratory procedures for reviewing patient results and notifying Health Care Providers that they serve. There was no U.S. distribution.
Root cause
Other
Status
Terminated
Product code
CGN
Quantity affected
2,521 kits
Distribution
Canada, Singapore, New Zealand, Germany, Hong Kong, Japan, and Australia
Date initiated
2005-03-11
Date posted
2005-04-28
Date terminated
2006-01-18
Location
Abbott Park, IL