ARCHITECT c16000 Processing Module, for In Vitro Diagnostics; List Number 3L77-01; Affected software: v2.10, v2.11, v2.1
FDA device recall Z-0147-2008 · posted 2007-11-07 · Terminated
Recall details
- Recalling firm
- Abbott Laboratories, Inc
- Reason for recall
- Incorrect Sample/Test Identification: Due to a software timing defect with certain versions of ARCHITECT cSystem Software, test results may be incorrectly associated to a Sample Identification during specific sample-processing conditions of the Robotic Sample Handler.
- Action taken
- Notification letters sent to consignees on 09/17/07. Consignees informed of the software defect and the conditions that would cause the software error to occur. Consignees told they would be contacted by an Abbott Field Service Rep who would install ARCHITECT System Software v3.11 and were asked to respond via fax using Customer Reply form acknowledging receipt and understanding of correction letter. Firm recalling to hospital/user level.
- Root cause
- Software change control
- Status
- Terminated
- Product code
- JJE
- Quantity affected
- 89 units.
- Distribution
- Worldwide Distribution: USA (including Puerto Rico), Argentina, Australia, Bahamas, Bermuda, Brazil, Cayman Islands, Canada, Chile, China, Colombia, Dominican Republic, Ecuador, El Salvador, Germany,
- Date initiated
- 2007-09-05
- Date posted
- 2007-11-07
- Date terminated
- 2010-10-04
- Location
- Irving, TX