Atlas FDA

ARCHITECT AUSAB Controls list number 1L82-10 have two bottles with the following components: 1) Negative Control and 2)

FDA device recall Z-1098-2007 · posted 2007-07-24 · Terminated

Recall details

Recalling firm
Abbott Diagnostic International, Ltd.
Reason for recall
During an internal study, the Architect AUSAB assay was found to read preparations of the WHO standard up to 53% higher than the WHO assigned value. This has the potential to cause elevated results and may impact the interpretation for a small percentage of results with grayzone or low-level reactive concentrations and higher proficiency sample results.
Action taken
All ARCHITECT AUSAB U.S. customers will be contacted. New customers will be contacted through a Product Information Letter that will be included in all ARCHITECT AUSAB Reagent Kits until assay is restandardized. All ARCHITECT AUSAB U.S. customers will be contacted. The following communication types were used: Product Correction Letter Customer Reply Form > Product Information Letter to be included in new kits to be distributed. Letters were sent to customers by Federal Express Priority mail. The Customers were instructed to: Please follow their laboratory procedures for evaluation of these results Disregard the automated interpretation of results for low-level reactive samples. A customer reply form will accompany the Product Correction letter. Effectiveness Check Depth: 100% of customers receiving the Product Correction letter. The start date of communication was May 14,2007.
Root cause
Other
Status
Terminated
Product code
LOM
Quantity affected
393
Distribution
Nationwide and Puerto Rico
Date initiated
2007-05-14
Date posted
2007-07-24
Date terminated
2011-06-10
Location
Barceloneta, PR

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.