AQUIOS CL Flow Cytometer, Catalog No. B30166; Intended for use with in vitro diagnostic flow cytometry applications usin
FDA device recall Z-2900-2016 · posted 2016-09-28 · Terminated
Recall details
- Recalling firm
- Beckman Coulter Inc.
- Reason for recall
- Beckman Coulter is recalling the AQUIOS CL System because the use of the reagent carousel can lead to misalignment of the carousel and damage to the prep probe.
- Action taken
- An Urgent Medical Device Recall letter dated 9/6/16 was sent to customers to inform them that the use of Positions 9 and 10 of the reagent carousel can lead to misalignment of the carousel and damage to the prep probe. The letter informs the customers of the impact of the recall, action to be taken, and the resolution by Beckman Coulter. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with any questions regarding this notice, are instructed to contact Customer Support Center: http://www.beckmancoulter.com/customersupport/support; call 1-800-369-0333 in the US and Canada; outside the US and Canada, contact their local Beckman Coulter Representative.
- Root cause
- Software Design Change
- Status
- Terminated
- Product code
- OYE
- Quantity affected
- 148 units total (12 units in US)
- Distribution
- US: IL, PA, NY, GA, WA, FL, SC. Affected geographies: Angola, Australia, Belgium, Botswana, Canada, Chile, Columbia, Czech Republic, Denmark, Finland, France, Germany, Ghana, Israel, Italy, Malaysia,
- Date initiated
- 2016-09-06
- Date posted
- 2016-09-28
- Date terminated
- 2017-08-07
- Location
- Brea, CA