AquaC UNO H [120V, 60Hz, 1000VA, US-Version - a reverse osmosis unit intended for use with hemodialysis systems. Freseni
FDA device recall Z-1769-2022 · posted 2022-09-19 · Open, Classified
Recall details
- Recalling firm
- Fresenius Medical Care Holdings, Inc.
- Reason for recall
- Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured
- Action taken
- Fresenius Medical Care Renal Therapies Group, LLC (FMCRTG) issued Urgent Medical Deice Correction letter (FA-2022-20-C) on 06/21/22 via First Class Mail. Letter states reason for recall, health risk and action to take: A Fresenius Medical Care Renal Therapies Group, LLC (FMCRTG) Water Systems Service Specialist will be contacting you to schedule a correction of your portable Water Treatment system. Only a Certified AquaC UNO H System Technician or FMCRTG Water Systems Service Specialist may perform this service. If choosing to correct via your own Certified AquaC UNO H System Technician, you will need to contact FMCRTG prior to performing the correction. FMCRTG will provide the necessary instruction via a Change Notification. For questions or concerns, please contact FMCRTG Technical Support Services at 888-553-1109.
- Root cause
- Under Investigation by firm
- Status
- Open, Classified
- Product code
- FIP
- Quantity affected
- 24 units
- Distribution
- US Nationwide distribution in the states of AZ, CA, DC, FL, GA, IN, MA, NV, NY, PA, TX.
- Date initiated
- 2022-06-21
- Date posted
- 2022-09-19
- Location
- Waltham, MA