Applied Medical, Separator Abdominal Access System REF: C0604 15 x 100mm Non-threaded Separator System with Universal Se

FDA device recall Z-0023-2009 · posted 2008-10-07 · Terminated

Recall details

Recalling firm
Applied Medical Resources Corp
Reason for recall
Potential inability to insufflate through the stopcock. Under certain circumstances, a flexible elastomeric component inside the 15mm trocar can stretch into a configuration that blocks the flow of insufflation gas (C02).
Action taken
Customers were notified via FedEx or UPS on June 25,2008 with an overnight letter. The letter advises that Applied Medical is conducting a voluntary recall of the 15mm Separator Access Systems, Models C0604, C0605, C0606 and C0607 due to the potential inability to insufflate through the stopcock. Per the letter, the firm is asking that all 15mm product listed in the letter be returned. A response form is provided.
Root cause
Process control
Status
Terminated
Product code
HET
Quantity affected
14,113 total
Distribution
Nationwide and Canada distribution.
Date initiated
2008-06-25
Date posted
2008-10-07
Date terminated
2012-03-26
Location
Rancho Santa Margarita, CA