Applied Medical's LAP CHOLE KIT ; Model Number: K0492 The device is used to close a clip in order to ligate a tubular st
FDA device recall Z-1507-2011 · posted 2011-03-03 · Terminated
Recall details
- Recalling firm
- Applied Medical Resources Corp
- Reason for recall
- Applied Medical is conducting a voluntary recall of specific lot numbers of the Direct Drive Laparoscopic Clip Applier, model number CA090, due to a potential defect in the jaws, possibly resulting in jaw breakage.
- Action taken
- Applied Medical sent an URGENT: MEDICAL DEVICE RECALL letter dated September 10, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to examine their inventory and complete the attached Voluntary Recall Notification Confirmation Form and fax it to 949-713-8832. Customers were asked to return all affected product. Secondary consignees were notified by Applied Medical on September 16, 2010. For any questions regarding this recall call 949-713-8041.
- Root cause
- Process design
- Status
- Terminated
- Product code
- GCJ
- Quantity affected
- 2,843 units for all products
- Distribution
- Worldwide Distribution - USA including AL, AR, AZ, CA, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OR, PA, SC, TN, TX, VT, WA, and WI and the countries of Australia
- Date initiated
- 2010-09-10
- Date posted
- 2011-03-03
- Date terminated
- 2011-03-09
- Location
- Rancho Santa Margarita, CA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.