Applied Medical's Inzii 12/15mm Retrieval System. Applieds Inzii Tissue Retrieval Bag, Model Number CD004, is classified
FDA device recall Z-0183-2014 · posted 2014-01-24 · Terminated
Recall details
- Recalling firm
- Applied Medical Resources Corp
- Reason for recall
- Applied Medical is conducting a voluntary recall of the Inzii 12/15mm retrieval system. During shipment, the retrieval system packaging has the potential to become punctured with small holes, which could compromise the sterile barrier. The likelihood of this situation to occur and result in patient harm is highly unlikely; however, out of an abundance of caution for patient safety and a commitmen
- Action taken
- Customer notifications of the recall began the week of 10/21/2013 via UPS overnight with letters requesting that all affected products be returned to Applied. The notification letter is dated October 21, 2013 and titled "URGENT: MEDICAL DEVICE RECALL". The notification was accompanied by a Recall Notification Confirmation Form. The letter informed customers of the recall by providing the product description with codes (Model and Lot Numbers). reason for the recall, actions to take and contact information.
- Root cause
- Package design/selection
- Status
- Terminated
- Product code
- GCJ
- Quantity affected
- 29,506 units
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Hong Kong, India, Ir
- Date initiated
- 2013-10-21
- Date posted
- 2014-01-24
- Date terminated
- 2014-09-15
- Location
- Rancho Santa Margarita, CA