Aplio MX Diagnostic Ultrasound System SSA-780A Used for routine diagnostic imaging

FDA device recall Z-0121-2014 · posted 2013-10-30 · Terminated

Recall details

Recalling firm
Toshiba American Medical Systems Inc
Reason for recall
The firm initiated this recall of certain Toshiba Diagnostic Ultrasound Systems because it has been found that under certain conditions, the temperature of the transducer head may increase in 2D+PW mode or 2D+CDI (POWER, ADF) +PW mode. When the system is used under both conditions listed below, the temperature of the transducer head connected to the system may increase in temperature and possibly
Action taken
Toshiba sent an Urgent Medical Device Correction letter dated August 5, 2013. along with a Reply Form via USPS return receipt mail. The letter identified the product, the problem, and the action to be taken by the customer. Until the software is corrected customers were instructed to change the preset settings as described in the letter. Customers were asked to complete and return the attached form and fax it to the toll free number. The form also can be sent via email to raffairs@tams.com Customers with questions were instructed to call 800-421-1968 or contact their local Toshiba Representative at 800-521-1968. For questions regarding this recall call 714-730-5000.
Root cause
Device Design
Status
Terminated
Product code
IYN
Quantity affected
427 units (1352 total)
Distribution
Nationwide Distribution including and US Territory Puerto Rico
Date initiated
2013-08-05
Date posted
2013-10-30
Date terminated
2016-06-28
Location
Tustin, CA