Atlas FDA

API Listeria (REF 10300) API Listeria is a standardized system for the identification of Listeria which uses miniaturize

FDA device recall Z-0842-2014 · posted 2014-01-27 · Terminated

Recall details

Recalling firm
Biomerieux France Chemin De L'Or
Reason for recall
bioMrieux identified a visual defect and activity issue on the ZYM B reagent (REF 70493); leading to a false negative results of some API biochemical tests using the ZYM B reagent.
Action taken
bioMerieux sent an Urgent Product Removal Notice letter dated November 15, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to ensure the letter and guidance were distributed and received by all appropriate personnel in their organization, customers were instructed to discontinue use and destroy ZYM (B 70493), API NH (10400) API Listeria (10300) with ZYM B ampules having a visual defect. Customers still in possession of some kits of ZYM B (70493-Lots: 1) without visual defect that is used with API 20 Strep (20600), API Coryne (20900), API Staph (20500) and API ZYM (25200), have the option to: a) Use the kits of ZYM B (70493) still in your inventory with the following mandatory requirement for the API Strep and API Coryne: perform a daily quality control with the ATCC strain described in the paragraph Reading and Interpretation of the package insert of each API strip prior to using the ZYM B ampules. b) Use the kits of ZYM B (70493) still in your inventory without any requirement with API Staph and API ZYM, or c) Discard all kits of ZYM B (70493) still in their inventory. If you are still in possession of some kits of API NH (10400-Lots: See Annex 1) without visual defect of ZYM B, you have the option to: a) Use the kits of API NH (10400) still in your inventory with the mandatory requirement: perform a daily quality control with the ATCC strain 49917 Haemophilus paraphrophilus described in the paragraph Use of the reagents of the package insert of API NH prior to using the ZYM B ampules or, b) Discard all the kits of API NH (10400) still in your inventory. 5. If you are still in possession of some kits of API Listeria (10300-Lots: See Annex 1) without visual defect of ZYM B, we ask you to use the API Listeria (10300) still in your inventory with the mandatory requirement: perform a daily quality control with the ATCC strain 33090 Listeria innocua as
Root cause
Under Investigation by firm
Status
Terminated
Quantity affected
39439 kits (includes ZYM B reagents, API Listeria * API NH kits)
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to U.S. Nationwide; Algerie, Argentina, Armenia, Australia, Austria, Azerbaidjian, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bolivia,
Date initiated
2013-11-08
Date posted
2014-01-27
Date terminated
2017-11-30
Location
Marcy L'Etoile, France

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.