Apellis Injection Kit 29g Injection Needle -single use intended to inject fluids into, or withdraw fluids from, parts of
FDA device recall Z-2653-2023 · posted 2023-09-29 · Open, Classified
Recall details
- Recalling firm
- Apellis Pharmaceuticals, Inc.
- Reason for recall
- 19-gauge x 1" inch filter needles showed appearances of a small fraction of non-preferred characteristics. A causal relationship has not been established between the structural variations in this 19-gauge filter needle and the rare events of retinal vasculitis
- Action taken
- Apellis issued Medical Device Field Correction letters via email on 8/22/23 to Healthcare Practitioners and Distributors/Specialty Distributors and Pharmacy/ Logistics. The communication to the Healthcare Practitioners also includes update on the rare events of retinal vasculitis with real world use of SYFOVRE (pegcetacoplan injection). A causal relationship has not been established between the 19-gauge needle and the rare events of retinal vasculitis in the real world. Letter states reason for recall, health risk and action to take: Actions to be taken by the Customer/Distributor: Please note that as part of this field action, you are requested to: " Confirm your availability and participation by sending back an email to acknowledge receipt of this message. " Immediately cease distribution of the below injection kit lots and place any remaining inventory of the lots On Hold. " Provide a summary of current on-hand inventory of the injection kit lots below. " Provide a distribution report of the injection kit lots including customer name/address, lot shipped, quantity shipped, and date. Please provide the Distribution report in a maximum time of four business hours. Apellis will also be recommending that practitioners discontinue use of, and dispose, any remaining injection kits that contain the specific 19-gauge filter needle and recommend they instead use the injection kits containing the 18-gauge filter needle already in distribution. Apellis is now exclusively distributing injection kits with the 18-gauge filter needle. If you need to place a supplemental order to replace quarantined lots, please place them immediately through your normal ordering process and Apellis will expedite shipping replacement injection kits./ Apellis will also be notifying physicians that if they do not have an injection kit with the 18-gauge filter needle, they may request a new kit at no cost by emailing info@apellis.com or contacting their Apellis Territory Business Mana
- Root cause
- Under Investigation by firm
- Status
- Open, Classified
- Product code
- FMI
- Quantity affected
- 61,943 units
- Distribution
- US Nationwide distribution.
- Date initiated
- 2023-08-22
- Date posted
- 2023-09-29
- Location
- Waltham, MA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|