Atlas FDA

Aortic Arch Cannula and Coronary Artery Perfusion Cannula with Ballon (part number NA-55X6). The Aortic Arch Cannulae ar

FDA device recall Z-2756-2016 · posted 2016-09-08 · Terminated

Recall details

Recalling firm
Sorin Group USA, Inc.
Reason for recall
Sorin Group USA, Inc. announces a voluntary field for the Aortic Arch Cannula and Coronary Artery Perfusion Cannula With Ballon because the tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of the tip is less than the requirement.
Action taken
LivaNova issued an Urgent Safety Alert letter via certified mail on August 12, 2016 to all affected customers. The letter instructed customers to check their inventory to determine if the Cannulae belongs to the product code and lots listed. If the customer does not have any of the products described in the communication, please complete the response form below and return it per the instructions indicated on the form; otherwise, please check the tip angle to determine if it is conforming to specifications. If specification is not met, please file a complaint through your normal process whether you decide to keep it and use it or return the product to Sorin Group; if you are not willing to use it, return the affected parts to Sorin Group USA. Please contact Customer Service 1-S00-650-2623 or email CustomerService@livanova.com to have a Return Material Authorization (RMA) issued. Complete the Customer Response Form attached to this letter to acknowledge receipt of this letter. Please return this form no later than September 15, 2016. For questions regarding this recall call 303-467-6306.
Root cause
Process control
Status
Terminated
Product code
DWF
Quantity affected
200 units worldwide
Distribution
Worldwide Distribution - US including IL, MO, TX. Internationally to Canada and Italy.
Date initiated
2016-08-12
Date posted
2016-09-08
Date terminated
2017-03-27
Location
Arvada, CO

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
AORTIC ARCH CANNULAS (K943934)Sorin Group Italia S.R.L.1995-06-02