Anspach MIA 16 (16cm Minimally Invasive Attachment), Rx Only MIA16 attachments are intended for use with Motor Systems f

FDA device recall Z-0779-2014 · posted 2014-01-16 · Terminated

Recall details

Recalling firm
The Anspach Effort, Inc.
Reason for recall
Product did not reach expected sterility assurance level
Action taken
The firm, Aspach, sent an "URGENT: MEDICAL DEVICE LABELING CORRECTION" letter dated May 13, 2013 to its consignees/customers. The letter described the product, problem and actions to be taken. The consignees/ customers were instructed that Sterrad 100S system should not be used on the MIA 16 attachment (the information only applies to only Sterrad sterilization of this devices and not to any other attachments.), and complete and return the Return Receipt Requested Customer Reply form via fax to: 1-800-327-6661 or email: ProductSupportTeam@synthes.com. If you distribute any of the products to other services or facilities, please forward this information as appropriate. Should you have any queries, please do not hesitated to contact Anspach Product Support at (800) 327-6887. Hours of Operation are 8:30am-5:30pm Eastern Standard Time or email: ProductSupportTeam@synthes.com.
Root cause
Unknown/Undetermined by firm
Status
Terminated
Product code
ERL
Quantity affected
183
Distribution
Nationwide Distribution: FL, IL, CA. IN, MI, CA, NJ, TX, NE, NC, OR, PA, TN, DC, UT, CO, MO, KY, MD, WV, NY, and MN.
Date initiated
2013-05-06
Date posted
2014-01-16
Date terminated
2016-04-18
Location
Palm Beach Gardens, FL