Ansell, Derma Prene Ultra, Powder-Free, Smooth Surgical Gloves, REF 8515, Size 7 1/2, 4 boxes of 50, Sterile, Distribute

FDA device recall Z-1586-2009 · posted 2009-07-24 · Terminated

Recall details

Recalling firm
Ansell Healthcare Products LLC
Reason for recall
Premature degradation which could result in tearing during use.
Action taken
The firm initiated the recall on May 1, 2009 by letter to Health Canada and by letter dated May 4, 2009 to its US and Canadian consignees. The letters described the affected product, issue and actions for customers. The recall was extended to the hospital and clinic level requesting that the product be returned. Direct questions to your Ansell Healthcare Products LLC representative.
Root cause
Process control
Status
Terminated
Product code
KGO
Quantity affected
560,000 total units--all lots.
Distribution
Nationwide Distribution and Canada.
Date initiated
2009-05-01
Date posted
2009-07-24
Date terminated
2010-04-12
Location
Dothan, AL

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
DERMA PRENE POWDER-FREE SYNTHETIC SURGICAL GLOVES (CHEMOTHERAPY USE) (K984558)Ansell Perry1999-02-26